GODIYMODULES Voltage Reference Module

GODIYMODULES Precision Voltage Reference Module User Manual

Modèle : Voltage Reference Module

Introduction

The GODIYMODULES Precision Voltage Reference Module is designed for accurate voltage and resistance calibration. It features low-temperature drift components and is powered via a Type-C 5V port, eliminating the need for a battery. This module provides five distinct voltage reference levels and five resistance reference levels, making it suitable for verifying the accuracy and linearity of digital multimeters.

Produit terminéview

GODIYMODULES Precision Voltage Reference Module, top-down view

Figure 1 : De haut en bas view of the GODIYMODULES Precision Voltage Reference Module. This image shows the overall layout of the module, including the Type-C power input, DIP switches for selecting voltage, and output terminals.

The module incorporates an EMI filter for a cleaner power supply. It includes a green LED to indicate power status. The compact design (85x55x25mm) makes it convenient for various applications.

Installation et connexion

  1. Alimentation: Connect a 5V Type-C power source (0.5A minimum) to the Type-C port on the module. The green LED will illuminate to indicate power.
  2. Connexion de sortie : Use appropriate test leads to connect your multimeter or device under test to the red (+) and black (-) output terminals on the module.
  3. Environnement: For optimal accuracy, especially when verifying 4 1/2 digit multimeters, ensure the working environment maintains a stable temperature (16-30℃) and humidity (RH≤70%). Avoid direct airflow from air conditioners, which can introduce errors.
GODIYMODULES Precision Voltage Reference Module with Type-C port and DIP switches

Figure 2 : Gros plan view of the module showing the Type-C power input and the DIP switches for selecting reference values.

Mode d'emploi

Sélection du volumetage Référence

The module provides five voltage reference levels: 1.24V, 2.5V, 5V, 7.5V, and 10V. These are selected using the DIP switches labeled 1 to 5. Each switch corresponds to a specific voltage niveau.

  • Commutateur 1 : 1.24V
  • Commutateur 2 : 2.5V
  • Commutateur 3 : 5V
  • Commutateur 4 : 7.5V
  • Commutateur 5 : 10V

Caution: Only one DIP switch for voltage selection should be in the "ON" position at any given time. Activating multiple voltage switches simultaneously is not permitted and may lead to incorrect readings or module damage.

Selecting Resistance Reference

The module provides five fixed resistance reference levels: 10Ω, 100Ω, 1KΩ, 10KΩ, and 100KΩ. These are available at dedicated terminals on the board. To measure a specific resistance, connect your multimeter to the corresponding resistance terminals.

GODIYMODULES Precision Voltage Reference Module showing resistance outputs

Figure 3 : View of the module highlighting the fixed resistance output terminals (10R, 100R, 1K, 10K, 100K).

Usage for Multimeter Calibration

  • 3 1/2 Digit Multimeters: The module can be used to verify the accuracy and linearity of handheld 3 1/2 digit digital multimeters.
  • 4 1/2 Digit Multimeters: For 4 1/2 digit digital multimeters, it is crucial to maintain the same temperature and humidity conditions as specified in the multimeter's data sheet for accurate verification.

Entretien

  • Nettoyage: Veillez à ce que le module reste propre et exempt de poussière et de débris. Utilisez un chiffon doux et sec pour le nettoyer.
  • Stockage: Store the module in a dry environment, away from extreme temperatures and humidity.
  • Manutention: Manipulez le module avec précaution afin d'éviter tout dommage physique aux composants ou aux connexions.

Dépannage

  • Aucun indicateur d'alimentation (LED verte éteinte) :
    • Ensure the Type-C power cable is securely connected.
    • Verify that the 5V Type-C power source is functional and provides at least 0.5A.
  • Vol incorrecttage Sortie :
    • Check that only one voltage selection DIP switch is in the "ON" position.
    • Ensure proper connection to the multimeter.
    • Verify environmental conditions (temperature, humidity) for sensitive measurements.
  • Lectures instables :
    • Operate in a low airflow environment to prevent temperature fluctuations.
    • Avoid direct cold wind from air conditioners.

Caractéristiques

FonctionnalitéSpécification
Nom du produitVoltage Reference Module
MarqueMODULES GODIY
Alimentation électriqueType-C 5V port, 0.5A
Consommation d'énergie à vide45 mA
Indication de puissanceLumière LED verte
Vol de référencetage Niveaux1.24V, 2.5V, 5V, 7.5V, 10V (5 levels)
Reference Resistance Levels10Ω, 100Ω, 1KΩ, 10KΩ, 100KΩ (5 levels)
Resistor Material0.1% Metal film copper feet & TCR: 25PPM
Dimensions (Clear Box Size)85 mm x 55 mm x 25 mm
Poids net34g
Environnement de travail16-30℃, RH≤70%

Ressources supplémentaires

For a visual demonstration of a voltage reference module (note: this video may feature a similar but not identical product), please refer to the video below:

Vidéo 1: Demonstration of a Voltage Reference Module. This video provides a general overview of how a voltage reference module might be used, including handling and basic operation. While the module shown may differ slightly from the GODIYMODULES model, it illustrates the general principles of such devices.

Garantie et assistance

Specific warranty information for the GODIYMODULES Precision Voltage Reference Module is not provided in the product details. For warranty claims or technical support, please contact your retailer or the manufacturer directly.

Documents connexes - Voltage Reference Module

Préview WRYTECH PDVS 2 Mini v2 Volume numérique de précisiontagManuel d'utilisation de la source
Manuel d'utilisation complet du voltmètre numérique de précision portatif WRYTECH PDVS 2 Mini v2tagCe manuel couvre les caractéristiques, les spécifications, le fonctionnement, les réglages, l'étalonnage et le dépannage de cet instrument polyvalent utilisé dans la recherche, le développement et l'industrie.
Préview Manuel d'utilisation du multimètre numérique ANENG M113 - Caractéristiques, spécifications et fonctionnement
Manuel d'utilisation complet du multimètre numérique ANENG M113, couvrant ses caractéristiques, les précautions de sécurité, les méthodes de mesure, les spécifications techniques, les indicateurs de précision et les procédures de maintenance.
Préview Manuel d'instructions du multimètre numérique de poche ANENG M113
Manuel d'instructions complet pour le multimètre numérique de poche ANENG M113, couvrant les précautions de sécurité, la description de l'instrument, les méthodes de mesure, les spécifications techniques et l'entretien.
Préview Manuel d'utilisation du multimètre numérique HABOTEST HT122 (type stylo)
Manuel d'utilisation complet du multimètre numérique à stylo HABOTEST HT122. Découvrez ses fonctions de sécurité, ses modes automatique et manuel, la détection de tension sans contact (NCV), la mesure de séquence de phases, ses caractéristiques techniques et son entretien.
Préview Manuel d'utilisation du multimètre numérique intelligent résidentiel UNI-T UT123/UT123D
Manuel d'utilisation complet pour les multimètres numériques intelligents résidentiels UNI-T UT123 et UT123D, couvrant les fonctionnalités, les consignes de sécurité, les procédures d'utilisation, les spécifications et la maintenance.
Préview Manuel d'utilisation du multimètre numérique Xhorse MK3
Manuel d'utilisation du multimètre numérique Xhorse MK3, couvrant ses caractéristiques, son fonctionnement, ses avertissements de sécurité et ses spécifications pour des mesures électriques précises.