RCYAGO DO9100

Compteur d'oxygène dissous avec liquide de remplissage d'électrode

Brand: RCYAGO | Model: DO9100

1. Introduction

The RCYAGO Dissolved Oxygen Meter is a high-precision instrument designed for accurate measurement of dissolved oxygen (DO) and temperature in various water bodies. It utilizes a polarographic sensor with a thermistor for automatic temperature compensation, ensuring reliable readings across different conditions. This manual provides comprehensive instructions for setup, operation, and maintenance of your DO meter.

RCYAGO Dissolved Oxygen Meter with probe

Figure 1: RCYAGO Dissolved Oxygen Meter and Probe

2. Contenu de la boîte

Veuillez vérifier le contenu du colis pour vous assurer que tous les éléments sont présents :

  • 1 x RCYAGO Dissolved Oxygen Meter (Model: DO9100)
  • 1 x Electrode Probe
  • 6 x Probe Protective Covers (Membrane Sleeves)
  • 1 x Electrode Filling Solution
  • 1 x Anaerobic Water Powder (for calibration)
  • 1 x Manuel d'utilisation

3. Installation

3.1 Installation de la batterie

The meter requires 3 x 1.5V button batteries. These are typically pre-installed or included in the top cap of the device. Ensure they are correctly oriented before use.

3.2 Connexion des électrodes

Connect the electrode probe to the main meter unit by screwing the metal knob securely. Ensure a firm connection to prevent inaccurate readings.

Diagram showing dimensions, LCD display, power button, lock/temperature switch key, mode/calibration key, and electrode interface.

Figure 2: Meter Components and Dimensions

3.3 Filling the Probe Protective Cover

The probe protective cover contains a permeable membrane. Before first use, or when replacing the membrane sleeve, fill it with the provided electrode filling solution. This solution is crucial for the sensor's proper function. Do not press on the membrane itself.

Close-up of the highly sensitive probe with multiple probe protective covers shown. The image highlights the detachable nature of the probe and the included covers.

Figure 3: Highly Sensitive Probe and Protective Covers

4. Fonctionnement

4.1 Mise sous/hors tension et affichage

Appuyez sur le bouton d'allumage () to turn the meter on. The digital screen will illuminate with a backlight, displaying DO readings and temperature. The backlight improves readability in low-light conditions. The device features an auto-power-off function after 8 minutes of inactivity to conserve battery life.

Close-up of the meter's HD backlight display showing readings, and an open battery compartment.

Figure 4: HD Backlight Display and Battery Compartment

4.2 Modes de mesure

The meter can display dissolved oxygen in mg/L (parts per million, ppm) or as a saturation percentage (%). Press the MODE/ÉTAL button to toggle between these modes.

4.3 Compensation de température

The meter features automatic temperature compensation (ATC) within the range of 0-50°C (32-122°F). This ensures that dissolved oxygen readings are accurate despite variations in water temperature, as oxygen solubility is affected by temperature.

Dissolved oxygen meter probe submerged in an aquarium, illustrating automatic temperature compensation.

Figure 5: Automatic Temperature Compensation in action

4.4 Étalonnage

Regular calibration is essential for accurate measurements. The meter supports two-point calibration: 100% air calibration and 0% anaerobic water calibration.

4.4.1 100% Air Calibration

  1. Ensure the probe is exposed to air.
  2. Appuyez longuement sur le MODE/ÉTAL bouton pendant 5 secondes.
  3. Release the button. The screen will flash 3 times at 100%. Calibration is complete.

4.4.2 0% Anaerobic Water Calibration

  1. Prepare anaerobic water: Pour 1 packet of anaerobic powder into the attached calibration bottle, add 30ml of purified water or tap water, and wait 5-10 minutes for the powder to fully dissolve.
  2. Submerge the electrode in the anaerobic water.
  3. Appuyez sur le MODE/ÉTAL key to switch to mg/L mode.
  4. Wait for the value to drop to 0.
  5. Appuyez longuement sur le MODE/ÉTAL bouton pendant 5 secondes.
  6. Release the button. The screen will flash 3 times at 0%. Calibration is complete.
Diagram illustrating the steps for 0% anaerobic water calibration and 100% air calibration.

Figure 6: Calibration Procedures

4.5 Prendre une mesure

Once calibrated, submerge the electrode probe into the water sample you wish to measure. The meter will display the dissolved oxygen and temperature readings. Allow a few moments for the readings to stabilize.

Person holding the dissolved oxygen meter, with the probe submerged in a pond with fish, showing a real-world application.

Figure 7: Measuring Dissolved Oxygen in a Pond

4.6 Changement d'unités de température

Pour basculer entre Celsius (°C) et Fahrenheit (°F), appuyez longuement sur MAINTIEN/TEMP bouton.

4.7 Understanding Oxygen Solubility

Oxygen solubility in water is influenced by temperature, atmospheric pressure, and salinity. The provided table illustrates the saturation concentration of oxygen at different temperatures:

Table showing saturation concentration of oxygen in mg/L and percentage at various temperatures in Celsius and Fahrenheit.

Figure 8: Oxygen Saturation Concentration at Different Temperatures

5. Entretien

5.1 Probe Protective Cover (Membrane Sleeve)

The film inside the protective cover cannot be taken apart. If test values become inaccurate even after calibration, the membrane sleeve needs to be replaced. The meter comes with 6 spare covers.

5.2 Nettoyage

The meter unit and probe should be cleaned with care. The product care instructions specify 'Hand Wash Only'. Avoid harsh chemicals or abrasive materials.

5.3 Stockage

When not in use, ensure the probe protective cover is properly filled with solution and securely attached to protect the sensor. Store the meter in a cool, dry place.

6. Dépannage

  • Lectures inexactes : Ensure the electrode is properly connected and the membrane sleeve is filled with the electrode filling fluid. Perform calibration regularly (both air and anaerobic water calibration) as described in Section 4.4.
  • Problèmes d'affichage : Check battery levels. If the display is dim or erratic, replace the batteries.
  • Pas de lecture : Verify the electrode is fully submerged in the sample and that the connection to the meter is secure.
  • Échec de l'étalonnage : Ensure the anaerobic water powder is fully dissolved for 0% calibration. For air calibration, ensure the probe is exposed to stable ambient air. If issues persist, the membrane sleeve might need replacement.

7. Spécifications

FonctionnalitéDétail
Dimensions du produit6.1 x 0.01 x 0.01 pouces
Numéro de modèle de l'articleDO9100
Poids12.56 onces
MarqueRCYAGO
Dossier spécialHaute précision
CouleurNoir
Matériau extérieurPlastique
Type d'affichageNumérique avec rétroéclairage
Plage d'oxygène dissous0.0 à 40.0 mg / L
Dissolved Oxygen Saturation0.0-300.0%
Plage de température0.0-40.0°C (32.0-104.0°F)
Compensation automatique de la température0-50°C
Source d'énergie3 x 1.5V Button Batteries
Arrêt automatique8 minutes

8. Vidéos produits

8.1 Digital Dissolved Oxygen Meter - Basic Operation & Calibration

Cette vidéo offre un aperçu rapideview of the RCYAGO Dissolved Oxygen Meter, demonstrating how to power it on, install the electrode, and perform both air and oxygen-free water calibrations. It highlights the key steps for initial setup and ensuring accurate readings.

8.2 Portable Dissolved Oxygen Meter - Detailed Review et utilisation

This comprehensive video offers a detailed review of the portable dissolved oxygen meter. It covers unboxing, battery installation, electrode setup, and practical demonstrations of measuring dissolved oxygen in different water samples (tap water vs. boiled water). The video also discusses the included accessories like membrane caps and electrolyte solution, and explains the temperature conversion chart.

9. Garantie et assistance

For warranty information, technical support, or to purchase replacement parts (such as probe protective covers or electrode filling solution), please contact RCYAGO customer service. Refer to the product packaging or the official RCYAGO website pour les coordonnées.

Documents connexes - DO9100

Préview Manuel d'instructions de l'analyseur d'oxygène dissous intelligent de type stylo DO-9100
Ce document fournit des instructions pour l'analyseur d'oxygène dissous intelligent de type stylo DO-9100, couvrant ses fonctionnalités, ses paramètres techniques, son fonctionnement, sa maintenance et son dépannage.
Préview Manuel d'utilisation de la minuterie d'eau double RCYAGO SGW07
Manuel d'utilisation complet du programmateur d'arrosage double RCYAGO SGW07. Découvrez le produit ci-dessous.view, spécifications, installation, programmes d'arrosage automatique, arrosage manuel, délai de pluie, dépannage et état de la batterie pour votre système d'irrigation de jardin.
Préview Oxymètre portable Yinmik DO9100, BLE9100 : Politique de mise en œuvre
Il s'agit d'un appareil portable pour les instruments portables Yinmik modèles DO9100 et BLE9100. Vérifiez les caractéristiques techniques, le calibrage, l'expédition, l'observation et la garantie.