1. Introduction

Thank you for choosing the CAMWAY HW300PRO Paint Thickness Gauge. This digital device is designed for precise measurement of coating thickness on various metallic surfaces. It is particularly useful for assessing paint thickness on automobiles, as well as other non-conductive coatings on ferrous metals and non-ferromagnetic coatings on non-ferrous metals. Please read this manual thoroughly before operation to ensure correct usage and optimal performance.

2. Produit terminéview et fonctionnalités

The CAMWAY HW300PRO is a compact and portable paint thickness gauge equipped with advanced features for accurate and convenient measurements.

Caractéristiques principales :

  • Mesure de haute précision : Accurately measures the thickness of non-conductive coatings on metallic surfaces and non-ferromagnetic coatings on ferromagnetic metals.
  • Multiple Data Recording: Stores average, maximum, minimum, and differential values. Supports up to 50 data points for statistical analysis.
  • Three Measurement Units: Switchable units between microns (um), millimeters (mm), and mils for versatile applications.
  • Écran HD rétroéclairé : High-definition screen with backlight ensures clear readability of data in various lighting conditions.
  • Arrêt automatique : L'appareil s'éteint automatiquement après 3 minutes d'inactivité afin de préserver la batterie.
  • Conception ergonomique : Features an anti-slip groove design for comfortable handling and a compact size for easy portability.
  • Large application : Suitable for automotive paint inspection, manufacturing, workshops, paint shops, and industrial fabrication.

Composants:

CAMWAY HW300PRO Paint Thickness Gauge with labeled buttons and display

Figure 2.1 : Disposition du dispositif. This image displays the CAMWAY HW300PRO Paint Thickness Gauge with its LCD display, control buttons, and probe. Key labels include "LCD Display", "Probe Mode Switch/Clear Statistics", "Unit Switch", "Switch Machine (Power Button)", "Zero Calibration, Switching Average/Minimum/Maximum/Number", and "Backlight Switch/Automatic Shutdown Switch".

Close-up of CAMWAY HW300PRO Paint Thickness Gauge showing HD backlit display, anti-slip groove design, and high precision probe

Figure 2.2: Detailed Features. Un gros plan view highlighting the HD backlit display, the anti-slip groove design on the side for better grip, and the high-precision measurement probe at the bottom of the device.

  • Écran LCD : Shows measurement readings, unit, battery status, and mode indicators.
  • Bouton d'alimentation (): Allume/éteint l'appareil.
  • F/NF Button: Switches between Ferrous (F), Non-Ferrous (N), and Automatic (AUTO) measurement modes. Also used to clear statistics.
  • um/mm/mils Button: Changes the measurement unit (microns, millimeters, mils).
  • Light/CAL Button: Toggles backlight on/off. Long press to enter calibration mode.
  • AUTO OFF Button: Toggles automatic shutdown function.
  • Sonde: The sensor at the bottom of the device for contacting the surface to be measured.
  • Cordon: For secure handling and portability.

3. Installation

3.1 Installation de la batterie

The device requires two AAA batteries. These are typically included with the product.

  1. Repérez le couvercle du compartiment à piles à l'arrière de l'appareil.
  2. Faites glisser ou déclipsez le couvercle pour ouvrir le compartiment.
  3. Insérez deux piles AAA en veillant à respecter la polarité (+/-) indiquée à l'intérieur du compartiment.
  4. Replacez soigneusement le couvercle du compartiment à piles.
Image showing two AAA batteries next to the CAMWAY HW300PRO device, illustrating battery installation for auto-off function.

Figure 3.1 : Installation de la batterie. The device uses two AAA batteries, shown here in relation to the device's automatic shutdown feature.

3.2 Mise sous/hors tension

  • To power on, press the Power button (L'écran s'illuminera.
  • Pour éteindre l'appareil, maintenez le bouton d'alimentation enfoncé () for a few seconds until the display turns off.
  • The device will automatically shut down after 3 minutes of inactivity if the auto-off function is enabled.

4. Mode d'emploi

4.1 Mesure de base

  1. Assurez-vous que l'appareil est sous tension.
  2. Select the desired measurement unit (um, mm, or mils) by pressing the "um/mm/mils" button.
  3. Select the appropriate measurement mode (AUTO, F, or N) by pressing the "F/NF" button.
    • VOITURE: Automatically detects the substrate type (ferrous or non-ferrous).
    • F (Ferrous): For measurements on magnetic substrates like iron or steel.
    • N (Non-Ferrous): For measurements on non-magnetic substrates like aluminum or copper.
  4. Placez la sonde fermement et perpendiculairement sur la surface à mesurer.
  5. La mesure s'affichera à l'écran.
  6. Lift the device from the surface to take another measurement.
Hand holding CAMWAY HW300PRO Paint Thickness Gauge against a car door handle, showing a measurement reading on the display.

Figure 4.1: Taking a Measurement. The image demonstrates how to position the gauge's probe on a car surface to obtain a thickness reading.

4.2 Fonction de rétroéclairage

To activate or deactivate the display backlight, press the "Light/CAL" button briefly. The backlight improves visibility in low-light conditions.

Comparison of CAMWAY HW300PRO display with backlight off and on, showing improved visibility.

Figure 4.2: Backlight Function. This image illustrates the difference in display visibility with the backlight off versus on, demonstrating enhanced readability.

5. Étalonnage

Regular calibration ensures the accuracy of your measurements. The device supports zero calibration and multi-point calibration.

5.1 Étalonnage du zéro

  1. Ensure the device is powered on and in the desired measurement mode (F or N).
  2. Place the probe directly onto the bare metal substrate (included standard aluminum or a known bare ferrous/non-ferrous surface).
  3. Press and hold the "Light/CAL" button until "ZERO CAL" appears on the display.
  4. Release the button. The device is now zero-calibrated for that specific substrate type.

5.2 Étalonnage multipoint

For enhanced accuracy, especially when measuring specific coating thicknesses, use the included calibration films.

  1. Perform a zero calibration first (Section 5.1).
  2. Place one of the calibration films (e.g., 0.05mm) onto the bare metal substrate.
  3. Place the probe firmly onto the calibration film.
  4. Adjust the reading on the device to match the known thickness of the film using the appropriate buttons (refer to the device's specific button functions for adjustment, typically the up/down arrows or mode buttons).
  5. Repeat this process for other calibration films (0.10mm, 0.25mm, 0.50mm, 1.00mm, 2.00mm) to cover the desired measurement range.
CAMWAY HW300PRO Paint Thickness Gauge with various calibration films (0.05mm, 0.10mm, 0.25mm, 0.50mm, 1.00mm, 2.00mm) and a standard aluminum substrate.

Figure 5.1: Calibration Components. This image shows the CAMWAY HW300PRO device along with the set of six calibration films and the standard aluminum substrate used for accurate calibration.

6. Gestion des données

The device can store and display various statistical data from your measurements.

  • ViewStatistiques d'utilisation : After taking multiple measurements, the device can display the average (AVG), maximum (MAX), minimum (MIN), and differential (DIF) values. Refer to the device's button layout (Figure 2.1) for specific button presses to cycle through these statistics.
  • Clearing Statistics: To clear the recorded data and start a new set of statistics, press the "F/NF" button (or the designated clear button if different) for a few seconds until the statistics reset. The device supports up to 50 data points for statistics.

7. Entretien

  • Nettoyage: Wipe the device with a soft, dry cloth. Do not use abrasive cleaners or solvents. Ensure the probe tip is clean and free of debris for accurate readings.
  • Stockage: Rangez l'appareil dans un endroit sec et frais, à l'abri de la lumière directe du soleil et des températures extrêmes. En cas de stockage prolongé, retirez les piles pour éviter les fuites.
  • Entretien de la sonde : The probe is a sensitive component. Avoid dropping the device or subjecting the probe to excessive force.

8. Dépannage

Problème Cause possible Solution
L'appareil ne s'allume pas. Piles faibles ou mortes ; installation incorrecte des piles. Remplacez les piles par des piles AAA neuves. Vérifiez la polarité des piles.
Lectures inexactes. Device not calibrated; dirty probe; incorrect measurement mode selected; surface not flat. Perform calibration (Section 5). Clean the probe tip. Ensure correct F/NF/AUTO mode. Measure on a flat surface.
L'écran est sombre ou illisible. Rétroéclairage désactivé ; batterie faible. Press the "Light/CAL" button to turn on backlight. Replace batteries.
L'appareil s'éteint inopinément. Auto-off function enabled; low battery. Disable auto-off function if continuous use is needed. Replace batteries.
Cannot switch units or modes. Buttons not pressed correctly; device frozen. Ensure a firm, brief press. Try powering off and on again.

9. Spécifications

Fonctionnalité Détail
Numéro de modèle CAMWT414USV
Marque CAMWAY
Plage de mesure 0-2000 um (microns)
Résolution 1 um
Unités de mesure um (microns), mm (millimeters), mils
Modes de mesure AUTO, F (Ferrous), N (Non-Ferrous)
Afficher LCD rétroéclairé HD
Stockage des données Up to 50 data points for statistics (AVG, MAX, MIN, DIF)
Arrêt automatique Après 3 minutes d'inactivité
Alimentation électrique Piles 2 x AAA (incluses)
Dimensions du colis 14.1 x 5.6 x 3.7 cm
Poids 100 grammes
Composants inclus Coating thickness gauge, lanyard, 6 calibration films (0.05/0.25/0.10/0.50/1.00/2.00 mm), substrate, instruction manual.
CAMWAY HW300PRO Paint Thickness Gauge with dimensions and key features listed: FE/NFE, 3 measurement units, high precision, showing multiple data, auto power off, backlight.

Figure 9.1: Device Dimensions and Features. This image provides a visual representation of the device's physical dimensions and reiterates its core features.

10. Garantie et assistance

Information regarding specific warranty terms and conditions for the CAMWAY HW300PRO Paint Thickness Gauge is not available in the provided product data. For warranty claims, technical support, or further assistance, please contact the seller or manufacturer directly.

Manufacturer: CAMWAY

Seller: CAMWAYUK (as per buybox_winner fulfillment details)